Hydac ETS 1700 Manual de usuario Pagina 9

  • Descarga
  • Añadir a mis manuales
  • Imprimir
  • Pagina
    / 15
  • Tabla de contenidos
  • MARCADORES
  • Valorado. / 5. Basado en revisión del cliente
Vista de pagina 8
Decimal places (Point)
(only when set to “RANG”=“FREE”)
Number of decimal places which the ETS 1700
displays.
0 ..
0.000
0.0
Lower display range (Range Low)
(only when set to “RANG”=“FREE”)
Lower display range limit
-995 ..
9995
0.0
Upper display range (Range High)
(only when set to “RANG”=“FREE”)
Upper display range limit
-995 ..
9995
measur-
ing range
final value
Analogue output (Output)
”MAMP”: The analogue output supplies a 4..20 mA
signal
”VOLT”: The analogue output supplies a 0..10 V
signal
MAMP/
VOLT
VOLT
Hidden basic settings menu (Hide)
”YES”: The basic settings menu is hidden and can
only be accessed if the keys for calling up
the menu are held down when the supply
voltage is switched on.
YES/ NO NO
Version number (Version)
Display of the current software version (for reference
only)
To end basic settings (End)
YES/ NO
9. Programming enable
The unit has 2 types of programming enable which must both be set to ”ON” to change the
settings. The software programming enable can be set or removed during operation. It provides
protection from unintentional alteration. A programming disable via the hardware programming
enable means that no change to the settings can be carried out during operation. This serves,
for example, as a safety function or as protection against unauthorised alterations.
9.1 Altering the software programming enable
Vista de pagina 8
1 2 3 4 5 6 7 8 9 10 11 12 13 14 15

Comentarios a estos manuales

Sin comentarios